FDA Adverse Event Injury Summary report: N

VASOVIEW HEMOPRO, OUS

MDR report key: 3213715 · Received July 3, 2013

Report

Report Number
2242352-2013-00636
Event Type
Injury
Date Received
July 3, 2013
Date of Event
June 17, 2013
Report Date
June 18, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE HEMOPRO JAW PEELED APART AND FELL INTO THE PT. THE PIECE WAS RETRIEVED WITH THE DEVICE THROUGH THE ORIGINAL INCISION. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304219 VASOVIEW HEMOPRO, OUS ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000-W 25065861

Patients

Seq Age Sex Outcome Treatment
1 NA Other