FDA Adverse Event
Injury
Summary report: N
ENDOTAK RELIANCE
MDR report key: 3213688
·
Received July 9, 2013
Report
- Report Number
- 2124215-2013-09371
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 13, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD HAD MICRODISLODGEMENT ACCOMPANIED BY HIGH THRESHOLDS AND LOW R WAVE. THIS RV LEAD WAS REPOSITIONED IN A BIT MORE SEPTAL POSITION. SUBSEQUENTLY, ALL LEAD MEASUREMENTS WERE NORMAL AFTER REVISION. THE RV LEAD REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312997 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0292 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| L| R | 0292| 4135| E162 |