FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 1000 DEFIBRILLATOR
MDR report key: 3213473
·
Received July 9, 2013
Report
- Report Number
- 3015876-2013-00588
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 12, 2013
- Report Date
- June 12, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K042404
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. IT WAS DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A BURNT FIELD EFFECT TRANSISTOR (FET), DESIGNATOR Q8, ON THE ANALOG PCB ASSEMBLY. THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE.
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED EVENT AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON WHEN THE ON BUTTON WAS PUSHED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312821 | LIFEPAK(R) 1000 DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 1000 |