FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 1000 DEFIBRILLATOR

MDR report key: 3213473 · Received July 9, 2013

Report

Report Number
3015876-2013-00588
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
June 12, 2013
Report Date
June 12, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K042404
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. IT WAS DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A BURNT FIELD EFFECT TRANSISTOR (FET), DESIGNATOR Q8, ON THE ANALOG PCB ASSEMBLY. THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED EVENT AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON WHEN THE ON BUTTON WAS PUSHED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.