FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3213197 · Received July 9, 2013

Report

Report Number
2531779-2013-09962
Event Type
Malfunction
Date Received
July 9, 2013
Report Date
June 23, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/21/2013 WITH THE FOLLOWING FINDINGS: NO DAMAGE WAS OBSERVED TO THE PUMP KEYPAD COVER. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE; ALL OTHER KEYPAD BUTTONS RESPONDED PROPERLY. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE CONTRAST AND UP ARROW KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE PUMP DISPLAY SCREEN APPEARED DIM AND DISCOLORED. WHEN REPLACED WITH A TEST DISPLAY, THE SCREEN FUNCTIONED PROPERLY.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT T HIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE UP BUTTON WAS RESPONDING ONLY AFTER MULTIPLE PRESSES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312964 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR