FDA Adverse Event Injury Summary report: N

INSIGNIA

MDR report key: 3213102 · Received July 9, 2013

Report

Report Number
2124215-2013-09705
Event Type
Injury
Date Received
July 9, 2013
Date of Event
June 1, 2013
Report Date
October 8, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS PACEMAKER PRESENTED TO THE EMERGENCY ROOM FOLLOWING A SYNCOPAL EPISODE. THE PATIENT PRESENTED WITH INTERMITTENT LONG PAUSES AND AN ESCAPE RHYTHM IN THE 30 BEATS PER MINUTE (BPM) RANGE. THE DEVICE WAS UNABLE TO BE INTERROGATED WITH A PROGRAMMER. IT WAS NOTED THAT THE PATIENT HAD BEEN LOST TO FOLLOW UP. THE DEVICE WAS SUCCESSFULLY EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313424 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1296

Patients

Seq Age Sex Outcome Treatment
1 92 YR Hospitalization| L| R 1226| 4261| MISMATCH| 1296