FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 3213053 · Received July 9, 2013

Report

Report Number
2124215-2013-09124
Event Type
Injury
Date Received
July 9, 2013
Date of Event
June 3, 2013
Report Date
June 28, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY DISCOVERED ABRASION ON THE LEAD LUMEN 700-705 MM FROM THE TERMINAL PIN. INSPECTION OF THE ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATIONS OF LOSS OF CAPTURE OR DISLODGEMENT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE. A REVISION PROCEDURE WAS PERFORMED DUE TO SUSPECTED LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS PRODUCT IS NO LONGER IN SERVICE.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314195 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization| L| R 4136| 0185| N119| 4542