TELIGEN
Report
- Report Number
- 2124215-2013-10148
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- June 14, 2013
- Report Date
- June 14, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0026-2014
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE DEVICE WAS NOT RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DISPLAYED A MESSAGE VOLTAGE TOO LOW FOR PROJECTED REMAINING CAPACITY. HOWEVER, REMAINING LONGEVITY STATES SEVEN YEARS. ADDITIONALLY, INTERROGATION DIFFICULTY WAS EXPERIENCED WAS EVENTUALLY SUCCESSFUL. THE DEVICE WAS EXPLANTED AND REPLACED WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312754 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | E110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Hospitalization| L | 5076| E110| 6942 |