FDA Adverse Event Malfunction Summary report: N

INCEPTA

MDR report key: 3213043 · Received July 9, 2013

Report

Report Number
2124215-2013-10311
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
June 3, 2013
Report Date
August 27, 2015
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DISCUSSED TROUBLESHOOTING WITH THE CALLER AND REQUESTED THE CALLER CALL BACK IF INTERROGATION IS NOT SUCCESSFUL. EFFORTS TO OBTAIN ADDITIONAL PRODUCT ISSUE DETAILS WERE UNSUCCESSFUL. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SUBSEQUENTLY EXPLANTED FOR NORMAL BATTERY DEPLETION AND RETURNED FOR TESTING. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT INTERROGATION OF THIS DEVICE WAS UNSUCCESSFUL. IT WAS NOTED THAT THERE WAS A LARGE BANDAGE ON THE WOUND WHICH COULD BE AFFECTING THE ABILITY TO INTERROGATE DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313539 INCEPTA IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND N164

Patients

Seq Age Sex Outcome Treatment
1 91 YR N164| 4470| 4525| H219| 4518