FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3213022 · Received July 9, 2013

Report

Report Number
2124215-2013-10642
Event Type
Injury
Date Received
July 9, 2013
Date of Event
May 13, 2011
Report Date
June 18, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT-OF-RANGE (OOR) PACING IMPEDANCES OF GREATER THAN 2000 OHMS. AS A RESULT, A PORTION OF THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313477 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0175

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L| R H219| 4517| 4087| 0175