FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 3212918 · Received July 9, 2013

Report

Report Number
2124215-2013-09863
Event Type
Injury
Date Received
July 9, 2013
Date of Event
June 6, 2013
Report Date
August 2, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

BOSTON SCIENTIFIC RECEIVED ADDITIONAL INFORMATION THAT THIS DEVICE WAS EXPLANTED AND REPLACED. THE DEVICE WAS LATER RETURNED FOR LABORATORY ANALYSIS. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT REMAINS IN SERVICE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED END OF LIFE (EOL) EARLY DUE TO CHARGE TIME (CT). BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) RECOMMENDED REMOVING THE DEVICE. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312774 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| L| R 4087| T167| 0185