FDA Adverse Event
Injury
Summary report: N
FLEXTEND
MDR report key: 3212911
·
Received July 9, 2013
Report
- Report Number
- 2124215-2013-09585
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 10, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE LEAD WAS SURGICALLY ABANDONED AND WILL NOT BE RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS, ELEVATED THRESHOLDS WITH NOISE AND OVERSENSING. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND REPLACED WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314225 | FLEXTEND | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 94 YR | Hospitalization| L | 4088| 1190 |