FDA Adverse Event Malfunction Summary report: N

5.0CM SHORT ATTACHMENT

MDR report key: 3212869 · Received July 9, 2013

Report

Report Number
1045834-2013-02694
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
June 13, 2013
Report Date
June 12, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE ACTUAL DEVICE WAS AVAILABLE FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. A VISUAL ASSESSMENT REVEALED THAT THE BEARINGS WERE DAMAGED. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. EVIDENCE INDICATES THIS WAS DUE TO NORMAL WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).

Additional Manufacturer Narrative · 1

THE ACTUAL/REPRESENTATIVE DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ATTACHMENT DEVICE HAD "BROKEN BEARINGS". IT WAS UNKNOWN TO THE REPORTER IF THE DEVICE WAS USED IN SURGERY. A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE AND NO DELAYS TO SURGERY WERE REPORTED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313487 5.0CM SHORT ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1