FDA Adverse Event Malfunction Summary report: N

UNIVERSAL CHUCK WITH T-HANDLE

MDR report key: 3212859 · Received July 9, 2013

Report

Report Number
8030965-2013-04257
Event Type
Malfunction
Date Received
July 9, 2013
Report Date
June 11, 2013
Manufacturer
SYNTHES GMBH
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. AN ADDITIONAL EVALUATION WAS COMPLETED: THE DEVICE WAS RECEIVED WITH THE HANDLE FALLING APART. THE INVESTIGATION HAVE SHOWN THAT THE COMPLETE UNIVERSAL CHUCK IS DISMANTLED. THIS IS LIKELY DUE TO REPROCESSING CARE MAINTENANCE. THE MANUAL PRE-CLEANING INCLUDES ULTRASONIC CLEANING; THIS ALLOWS EFFECTIVE CLEANING AND REPROCESSING ARE GUARANTEED WITH STEAM STERILIZATION. THIS DEVICE IS OVER 13 YEARS OLD (IT WAS MADE IN MARCH 2000) AND IT CANNOT BE DETERMINED HOW THE DEVICE WAS CONTAMINATED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: REPORTEDLY THE HANDLE WAS DEFECTIVE AND FELL APART DURING A PROCEDURE ON AN UNKNOWN DATE, CAUSING IT TO BE CONTAMINATED BY DIRT AND MUD, AND WAS VERY CRITICAL. REPORTEDLY IN THE PREPARATION INFORMATION THERE IS NOT ANY NOTE THAT THIS INSTRUMENT SHOULD BE DISASSEMBLED BEFORE CLEANING AND THE FACILITY FELT THAT THIS INSTRUMENT WOULD ALWAYS BE CONTAMINATED AS A RESULT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314339 UNIVERSAL CHUCK WITH T-HANDLE MISC ORTHO SURGICAL INSTR LXH SYNTHES GMBH 1053304

Patients

Seq Age Sex Outcome Treatment
1