FDA Adverse Event
Malfunction
Summary report: N
Z-800 INFUSION PUMP
MDR report key: 3212820
·
Received July 2, 2013
Report
- Report Number
- MW5030795
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- January 14, 2013
- Report Date
- March 5, 2013
- Manufacturer
- ZYNO MEDICAL LLC
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MI, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PUMP USED FOR 1 PT MOST OF DAY WITH HYDRATION BEFORE AND AFTER CHEMO. PUMP WAS PROGRAMMED FOR RATE OF 600 CC/HR. INFUSION WAS COMPLETED BUT PUMP CONTINUED TO RUN EVEN WITH TRYING THE STOP/START BUTTON, OFF/ON BUTTON AND UNPLUGGING MACHINE. DOOR WAS OPENED AND TUBING REMOVED - NO ALARMS ACTIVATED AND MACHINE CONTINUED THE PUMPING MODE. SEVERAL PEOPLE ATTEMPTED TO STOP MACHINE WITHOUT SUCCESS AND IT WAS STILL RUNNING, UNPLUGGED WHEN WE LEFT WORK THAT EVENING. NO LONGER WAS RUNNING UPON RETURN IN AM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 301920 | Z-800 INFUSION PUMP | Z-800 INFUSION PUMP | FRN | ZYNO MEDICAL LLC | Z 800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |