FDA Adverse Event Malfunction Summary report: N

Z-800 INFUSION PUMP

MDR report key: 3212820 · Received July 2, 2013

Report

Report Number
MW5030795
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
January 14, 2013
Report Date
March 5, 2013
Manufacturer
ZYNO MEDICAL LLC
Product Code
FRN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PUMP USED FOR 1 PT MOST OF DAY WITH HYDRATION BEFORE AND AFTER CHEMO. PUMP WAS PROGRAMMED FOR RATE OF 600 CC/HR. INFUSION WAS COMPLETED BUT PUMP CONTINUED TO RUN EVEN WITH TRYING THE STOP/START BUTTON, OFF/ON BUTTON AND UNPLUGGING MACHINE. DOOR WAS OPENED AND TUBING REMOVED - NO ALARMS ACTIVATED AND MACHINE CONTINUED THE PUMPING MODE. SEVERAL PEOPLE ATTEMPTED TO STOP MACHINE WITHOUT SUCCESS AND IT WAS STILL RUNNING, UNPLUGGED WHEN WE LEFT WORK THAT EVENING. NO LONGER WAS RUNNING UPON RETURN IN AM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301920 Z-800 INFUSION PUMP Z-800 INFUSION PUMP FRN ZYNO MEDICAL LLC Z 800

Patients

Seq Age Sex Outcome Treatment
1