FDA Adverse Event Death Summary report: N

SIGMA 300 DR

MDR report key: 3212715 · Received July 9, 2013

Report

Report Number
3004209178-2013-11477
Event Type
Death
Date Received
July 9, 2013
Report Date
July 18, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION RECEIVED INDICATED THE CAUSE OF DEATH AS CONGESTIVE HEART FAILURE, AORTIC STENOSIS AND DEMENTIA. FURTHER REVIEW PROMPTED A CHANGE IN THE DEVICE ANALYSIS RESULTS. THE CHANGE IS REFLECTED IN THIS REPORT UNDER - CONCLUSION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. CONCOMITANT PRODUCTS: 4058M45 IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 1995; 5024M-58 IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 1995. (B)(4).

Description of Event or Problem · 1

THE IMPLANTABLE PULSE GENERATOR (IPG) WAS RETURNED WITH NO INFORMATION, ANALYZED AND TESTED OUT OF SPECIFICATION. FURTHER REVIEW OF THE MANUFACTURER¿S DATABASE INDICATED THE PATIENT IS DECEASED AND DIED APPROXIMATELY EIGHT YEARS POST IMPLANT. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313336 SIGMA 300 DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO SDR303B

Patients

Seq Age Sex Outcome Treatment
1 00096 YR Death