TELIGEN
Report
- Report Number
- 2124215-2013-10561
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- May 13, 2013
- Report Date
- October 8, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0026-2014
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. A FAULT CODE 1003 WAS CONFIRMED TO HAVE BEEN RECORDED ON (B)(6) 2013. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH DEVICE PERFORMANCE WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. AFTER THE DEVICE CASE WAS OPENED, MICROSCOPIC VISUAL INSPECTION DID NOT REVEAL ANY IRREGULARITIES. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED (2.949 VOLTS), BUT STILL SUPPORTED FULL DEVICE FUNCTION. THE BATTERY WAS THEN REMOVED, SO THAT AN EXTERNAL POWER SUPPLY COULD BE CONNECTED TO THE DEVICE FOR FURTHER ANALYSIS. A HIGHER THAN NORMAL CURRENT USAGE WAS OBSERVED. ELECTRICAL TESTING ISOLATED THE HIGH CURRENT CONDITION TO TWO LOW-VOLTAGE CAPACITORS CONNECTED TO THE DEVICE¿S BATTERY. LOW-VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH-VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT UPON INTERROGATION OF THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) A FAULT CODE 1003 - VOLTAGE TOO LOW FOR PROJECTED REMAINING CAPACITY WAS OBSERVED. THE HEALTH CARE PROFESSIONAL (HCP) CONSULTED WITH BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) AND IT WAS DISCUSSED THIS WAS AN EARLY WARNING OF PREMATURE BATTERY DEPLETION, DESPITE THE DEVICE INDICATING 9.5 YEARS REMAINING. A SAVE TO DISK WAS DONE AND EVALUATED BY A BOSTON SCIENTIFIC ENGINEER. PRESENTLY THERAPY WAS UNAFFECTED, AS THERE WAS RESERVE CAPACITY, HOWEVER, THERE WAS A HIGHER CURRENT DRAIN THAN EXPECTED. IT WAS RECOMMENDED THE DEVICE BE CHANGED OUT. BASED ON THE DATA, THERE WAS RESERVE TO MAINTAIN NORMAL THERAPY AND FUNCTION FOR TWENTY-EIGHT DAYS. SUBSEQUENTLY, SURGICAL INTERVENTION WAS PERFORMED AND THE DEVICE WAS SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312791 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization| L| R | E102| 0185 |