FDA Adverse Event Injury Summary report: N

CONTAK RENEWAL

MDR report key: 3212640 · Received July 9, 2013

Report

Report Number
2124215-2013-09035
Event Type
Injury
Date Received
July 9, 2013
Date of Event
May 24, 2013
Report Date
June 11, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

DETAILED ANALYSIS IS BEING DONE ON THIS DEVICE. ONCE ANALYSIS IS COMPLETE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ENTIRE SYSTEM WAS EXPLANTED FOR INFECTION. DURING THE EXPLANT PROCEDURE, IT WAS NOTED THAT THE DEVICE SETSCREW WAS STUCK. THE DEVICE AND LEADS WERE EXPLANTED. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314205 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H217

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| L| R 4554| 0184| 1280| H217| 4470| 4469