INSIGNIA
Report
- Report Number
- 2124215-2013-10027
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- April 17, 2013
- Report Date
- October 8, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE BATTERY STATUS WAS ERT. THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY (EVEN SLIGHT) CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED ERT EARLIER THAN PREVIOUSLY ESTIMATED.
RECORDS INDICATE THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER HAD REACHED ELECTIVE REPLACEMENT TIME (ERT). A COUPLE MONTHS PRIOR THE DEVICE DISPLAYED A REMAINING LONGEVITY OF ONE YEAR. NO OTHER DEVICE ANOMALIES WERE NOTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314169 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 94 YR | Hospitalization| L| R | 4054| 4053| 1290 |