TELIGEN
Report
- Report Number
- 2124215-2013-08977
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- May 27, 2013
- Report Date
- June 27, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0026-2014
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
(B)(4).
UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED TWO LOW VOLTAGE BATTERY FAULTS (FAULT CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. INITIAL AUTOMATED TESTING VERIFIED BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE. AFTER THE TITANIUM CASE WAS OPENED, MICROSCOPIC VISUAL INSPECTION DID NOT REVEAL ANY IRREGULARITIES. THE BATTERY WAS THEN REMOVED, SO THAT AN EXTERNAL POWER SUPPLY COULD BE CONNECTED TO THE DEVICE FOR FURTHER ANALYSIS. A HIGHER THAN NORMAL CURRENT USAGE WAS OBSERVED. ELECTRICAL TESTING ISOLATED THE HIGH CURRENT CONDITION TO TWO COMPROMISED LOW VOLTAGE CAPACITORS THAT ARE CONNECTED TO THE DEVICE¿S BATTERY. THIS TYPE OF COMPONENT MALFUNCTION CAN RESULT IN A HIGH CURRENT DRAIN, WHICH OVER TIME CAN RESULT IN PREMATURE BATTERY DEPLETION, AS WAS OBSERVED IN THIS CASE.
ENGINEERING REVIEW OF THE DEVICE DATA CONFIRMED A LOW-VOLTAGE FAULT (FAULT CODE 1003) HAD OCCURRED. THERE WERE NO OTHER FAULTS OR RESETS STORED WITHIN DEVICE MEMORY. THE VOLTAGE WAS 3.020 VOLTS, CONFIRMING THERAPY DELIVERY WAS UNAFFECTED. THE DEVICE WAS MALFUNCTIONING SUCH THAT THE BATTERY STATUS INDICATORS WERE NOT REFLECTING THE DEPLETION CONDITION AND WERE THEREFORE INACCURATE. A REVIEW OF THE BATTERY VOLTAGE MEASUREMENT DATA INDICATED THE CURRENT DRAIN WAS CONSISTENT OVER TIME, SO THERE WAS SUFFICIENT RESERVE FOR THE DEVICE TO MAINTAIN NORMAL THERAPY FUNCTIONS FOR APPROXIMATELY 28 DAYS. TECHNICAL SERVICE'S RECOMMENDATION FOR DEVICE REPLACEMENT WAS CONFIRMED BY ENGINEERING'S REVIEW OF THE DEVICE MEMORY. THE DEVICE WAS THEN EXPLANTED AND REPLACED. THIS REPORT WILL BE UPDATED WHEN THE DEVICE IS RETURNED AND ANALYSIS IS COMPLETE.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REPORTED HEARING BEEPING TONES FROM THE DEVICE. THE DEVICE WAS INTERROGATED AND A FAULT CODE 1003 WAS DISPLAYED. A BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DISCUSSED THE FAULT CODE AND RECOMMENDED DEVICE REPLACEMENT. IT WAS ALSO REQUESTED THAT A COPY OF THE DEVICE MEMORY BE SAVED TO A DISK FOR ENGINEERING EVALUATION. THE FIELD REPRESENTATIVE REPORTED THAT THE DEVICE WAS SCHEDULED FOR REPLACEMENT THE FOLLOWING DAY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312214 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | F110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |