ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-08447
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- April 3, 2013
- Report Date
- May 17, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS AVAILABLE AT THIS TIME, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION LESS THAN TWO MONTHS AGO, THIS PATIENT CAME TO THE HOSPITAL IN THE MORNING FOLLOWING THE PREVIOUS EVENING WHERE THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SHOCKED HIM. HE ALSO REPORTED PREVIOUSLY THE ICD EMITTED BEEP TONES. UPON INTERROGATION AT THE HOSPITAL, THE DEVICE REVERTED TO A RESET SAFETY MODE; HOWEVER, IT REMAINED UNCLEAR WHICH TYPE OF RESET HAD OCCURRED. THE FOLLOWING DAY, THE ICD WAS EXPLANTED, REPLACED, AND RETURNED FOR ANALYSIS. THE CHRONIC LEADS WERE RECONNECTED TO THE NEW DEVICE AND TESTED SUCCESSFULLY. NO ADDITIONAL ADVERSE EFFECTS WERE REPORTED. THE ICD WAS SUBSEQUENTLY RETURNED ONE AND ONE HALF MONTHS LATER FOR ANALYSIS. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION NOTED AN ARC MARK ON THE DEVICE CASE. ADDITIONALLY, ANALYSIS CONFIRMED AN INABILITY TO COMMUNICATE WITH THE DEVICE. ELECTRICAL TESTING PERFORMED INDICATED A BATTERY VOLTAGE OF 1.1V, AND VERIFIED THAT THE SHOCKING OUTPUT BRIDGE WAS INTACT. HOWEVER, AS THE ARCING ON THE DEVICE CASE WAS DIRECTLY OVER THE DEVICE CIRCUITRY, THE ARCING CAUSED INTERNAL DEVICE DAMAGE AND RESULTED IN THE OBSERVED SAFETY MODE OPERATION. BOSTON SCIENTIFIC'S HISTORICAL TREND DATA INDICATES THAT THE CAUSE OF ARCING DAMAGE LIKE THAT SEEN WITH THIS DEVICE IS AN EXTERNAL HIGH VOLTAGE SHORT WHICH MAY CAUSE INTERNAL DAMAGE TO THE CIRCUITRY. THE DAMAGE TO THE DEVICE WAS MOST LIKELY DUE TO A SHORTED LEAD CONDITION THAT OCCURRED DURING DELIVERY OF A SHOCK. DEVICES ARE NOT EXPECTED TO PERFORM TO SPECIFICATION POST-ARCING EVENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314409 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | F110| MISMATCH| T165| 0148 |