FDA Adverse Event
Malfunction
Summary report: N
ENDOTAK RELIANCE
MDR report key: 3212497
·
Received July 9, 2013
Report
- Report Number
- 2124215-2013-09289
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- May 20, 2013
- Report Date
- May 20, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED PACING INHIBITION DUE TO OVERSENSING OF ATRIAL ACTIVITY. THE PATIENT IS PACER DEPENDENT AND EXPERIENCED ASYSTOLE GREATER THAN TWO SECONDS. THE SENSITIVITY WAS INCREASED AND THE LEAD REMAINS IMPLANTED. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 313043 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | 0157| A135| 4087| E110 |