FDA Adverse Event Malfunction Summary report: N

RESOLUTION CLIP CLIPPING DEVICE

MDR report key: 3212491 · Received July 9, 2013

Report

Report Number
3005099803-2013-05633
Event Type
Malfunction
Date Received
July 9, 2013
Date of Event
April 22, 2013
Report Date
June 12, 2013
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MND
PMA / PMN Number
K040148
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). (B)(4) FOR THE INVESTIGATION FINDING OF BUSHING BENT DUE TO CLIP MASHED ONTO THE BUSHING. A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE CLIP ASSEMBLY WAS MASHED ONTO THE BUSHING AND THE CLIP ASSEMBLY COULD NOT BE DEPLOYED. THE PRONGS COULD NOT BE OPENED OR CLOSED BUT WERE NOT LOCKED INTO THE CAPSULE. THERE WAS A KINK IN THE CONTROL WIRE. THE OVERSHEATH ASSEMBLY WAS NOT RETURNED. BASED ON THE CONDITION OF THE RETURNED DEVICE, THE REPORTED FAILURE COULD NOT BE CONFIRMED. HOWEVER, IT IS LIKELY THAT THE DIFFICULTIES EXPERIENCED DURING DEPLOYMENT WERE DUE TO ANATOMICAL OR PROCEDURAL FACTORS SUCH AS TORTUOUS ANATOMY OR MANEUVERING OF THE DEVICE. THEREFORE, A REVIEW AND ANALYSIS OF ALL AVAILABLE INFORMATION INDICATED THAT THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USED TO STOP A BLEED IN THE STOMACH ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP WAS UNABLE TO BE OPENED. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER RESOLUTION CLIP. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED AS BEING STABLE. NOTE: INVESTIGATION RESULTS REVEALED THE CLIP WAS MASHED ONTO THE BUSHING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313204 RESOLUTION CLIP CLIPPING DEVICE LIGATOR, ESOPHAGEAL MND BOSTON SCIENTIFIC - MARLBOROUGH M00522610 ML000466C3

Patients

Seq Age Sex Outcome Treatment
1