FDA Adverse Event Injury Summary report: N

CONTAK RENEWAL

MDR report key: 3212415 · Received July 9, 2013

Report

Report Number
2124215-2013-09461
Event Type
Injury
Date Received
July 9, 2013
Date of Event
May 18, 2013
Report Date
June 17, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CHARGE TIMES GREATER THAN THE CHARGE TIME LIMIT. END OF LIFE (EOL) WAS DECLARED FOLLOWING A SINGLE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE AND THE RESULTS INDICATED THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. IT WAS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, THE ERI TO EOL TIME PERIOD WAS SHORTENED DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE. THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS, AS THE BATTERY ITSELF HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

UPON RECEIPT IN THE POST MARKET QUALITY ASSUARANCE LABORATORY, INITIAL TESTING IDENFIIED A POTENTIAL INDICATION OF AN OUT OF SPECIFICATION CONDITION. THIS PRODUCT ISSUE WILL BE UPDATED WHEN EVALUATION IS COMPLETE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE BATTERY HAS REACHED END OF LIFE (EOL). THE DEVICE WAS EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314077 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H177

Patients

Seq Age Sex Outcome Treatment
1 88 YR Hospitalization H177| MISMATCH| 6024| 0010| 6931