CONTAK RENEWAL
Report
- Report Number
- 2124215-2013-09461
- Event Type
- Injury
- Date Received
- July 9, 2013
- Date of Event
- May 18, 2013
- Report Date
- June 17, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CHARGE TIMES GREATER THAN THE CHARGE TIME LIMIT. END OF LIFE (EOL) WAS DECLARED FOLLOWING A SINGLE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE AND THE RESULTS INDICATED THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. IT WAS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, THE ERI TO EOL TIME PERIOD WAS SHORTENED DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE. THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS, AS THE BATTERY ITSELF HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
UPON RECEIPT IN THE POST MARKET QUALITY ASSUARANCE LABORATORY, INITIAL TESTING IDENFIIED A POTENTIAL INDICATION OF AN OUT OF SPECIFICATION CONDITION. THIS PRODUCT ISSUE WILL BE UPDATED WHEN EVALUATION IS COMPLETE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE BATTERY HAS REACHED END OF LIFE (EOL). THE DEVICE WAS EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314077 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Hospitalization | H177| MISMATCH| 6024| 0010| 6931 |