ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09920
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Report Date
- June 11, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: REVIEW OF THE BLACK BOX DATA AND ALARM HISTORY REVEALED MULTIPLE ¿LOW CARTRIDGE¿ WARNINGS RECORDED. ON INVESTIGATION, THE PUMP RESPONDED APPROPRIATELY TO A LOW CARTRIDGE AND GAVE THE EXPECTED AUDIBLE AND VISUAL ALERT. THE KEYPAD WAS NOTED TO BE INTACT WITHOUT DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND DID NOT REVEAL ANY EVIDENCE OF DAMAGE, DEFECT OR CONTAMINATION OF THE BUTTON CONTACTS. INVESTIGATION WAS UNABLE TO DUPLICATE THE ALLEGED ISSUE OF THE PUMP NOT ALARMING WITH THE INSULIN IN THE CARTRIDGE IS LOW OR THE KEYPAD BUTTONS BEING UNRESPONSIVE. THE PUMP WAS FOUND TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER ALLEGED THAT BEGINNING A FEW MONTHS PRIOR, THE UP ARROW AND DOWN BUTTONS HAVE DELAYED RESPONSE WHEN PRESSED. THE REPORTER ALSO ALLEGED THAT THE PUMP DOES NOT ALARM WHEN THE INSULIN IN THE CARTRIDGE IS LOW. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED KEYPAD MALFUNCTION AND ALARM MALFUNCTION REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312275 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |