DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00320
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 7, 2013
- Report Date
- June 10, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- MLM
- PMA / PMN Number
- K060502
- Removal / Correction Number
- 2517506-11-17-2013-018-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
NO ADVERSE EVENT OCCURRED. THE CUSTOMER WAS DISSATISFIED WITH THE PERFORMANCE OF THE REAGENT. HOWEVER, THE CAUSE OF THE BIASED LOW RESULTS RELATIVE TO THE ALTERNATE METHODOLOGY IS UNDER INVESTIGATION BY SIEMENS HEALTHCARE DIAGNOSTICS INC. SIEMENS HAS CONFIRMED THAT THE DIMENSION TACR FLEX REAGENT CARTRIDGE LOT BB4087 AND ITS LINKED CALIBRATOR LOT 3BD029 MAY EXHIBIT LOW PATIENT SAMPLE RECOVERY ALTHOUGH QUALITY CONTROL (QC) MATERIALS HAVE NOT EXHIBITED RECOVERY OUTSIDE OF THE EXPECTED RANGE. SIEMENS INTERNAL COMPLAINT TESTING OF THE DIMENSION TACR FLEX REAGENT CARTRIDGE LOT BB4087 WITH THE LINKED CALIBRATOR LOT 3BD029 VERSUS THE REFERENCE LC/MS METHOD CONFIRMED AN AVERAGE BIAS OF -1.6 NG/ML [-2.1 NMOL/L], ACROSS THE ASSAY RANGE. INDIVIDUAL PATIENTS, PARTICULARLY SAMPLES WITH LOWER TACR VALUES (< 5 NG/ML [6.5 NMOL/L]), MAY SHOW MORE BIAS DUE TO THE TYPICAL PRECISION OF EACH METHODOLOGY. SIEMENS OBSERVED SOME LOW PATIENT LCMS TACROLIMUS SAMPLES (< 5 NG/ML [6.5 NMOL/L]) THAT GENERATED RESULTS BELOW THE DIMENSION TACROLIMUS ASSAY RANGE (< 1.2 NG/ML [1.6 NMOL/L]).
ORIGINAL MDR WAS FILED 2013-07-09. SIEMENS HEALTHCARE DIAGNOSTICS INC. ISSUED AN URGENT MEDICAL DEVICE CORRECTION LETTER, 13-48, IN JULY 2013. THE LETTER STATED THAT SIEMENS HAS CONFIRMED THAT THE DIMENSION® TACR FLEX REAGENT CARTRIDGE LOT BB4087 AND ITS LINKED CALIBRATOR LOT 3BD029 MAY EXHIBIT LOW PATIENT SAMPLE RECOVERY ALTHOUGH QUALITY CONTROL (QC) MATERIALS HAVE NOT EXHIBITED RECOVERY OUTSIDE OF THE EXPECTED RANGE. CUSTOMERS WERE INSTRUCTED TO DISCONTINUE USE AND IMMEDIATELY DISCARD ANY REMAINING INVENTORY OF DIMENSION® TACR FLEX® LOT BB4087 AND CALIBRATOR LOT 3BD029. SIEMENS IS INVESTIGATING THE ISSUE AND WILL IMPLEMENT ADDITIONAL CONTROLS TO ENSURE ACCEPTABLE BIAS PRIOR TO ANY FUTURE SHIPMENTS.
BIASED LOW RESULTS WERE OBTAINED ON TACROLIMUS (TACR) PATIENT SAMPLES IN A PATIENT CROSSOVER STUDY VERSUS AN ALTERNATE METHODOLOGY. THE REAGENT LOT WAS NOT IN USE AND PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIANS. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE BIASED LOW RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE BIASED LOW TACROLIMUS RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312269 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | TACROLIMUS (TACR) FLEX® REAGENT CARTRIDGE | MLM | SIEMENS HEALTHCARE DIAGNOSTICS INC | BB4087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |