ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09915
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Report Date
- June 11, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 08/29/2012 DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/08/2012WITH THE FOLLOWING FINDINGS:THE KEYPAD WAS FOUND TO BE INTACT. THE KEYPAD BUTTONS WERE FOUND TO BE RESPONSIVE. THE KEYPAD COVER WAS REMOVED; CONTAMINATION WAS FOUND UNDER THE BUTTON KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED DISPLAY SCREEN. A NEW TEST SCREEN WAS INSERTED; THE DISPLAY SCREEN WAS FOUND TO BE FUNCTIONING APPROPRIATELY AND ILLUMINATED TO NORMAL CONTRAST.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE; ALL BUTTONS DIFFICULT TO PUSH. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312172 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR |