STERRAD NX STERILIZER
Report
- Report Number
- 2084725-2013-00310
- Event Type
- Malfunction
- Date Received
- July 9, 2013
- Date of Event
- June 28, 2013
- Report Date
- June 28, 2013
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K042116
- Removal / Correction Number
- Z-0744-0746-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. THE VACUUM PUMP AND VACUUM PUMP OIL WAS REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE ON (B)(6) 2013.
ADDITIONAL SERVICE PROVIDED - FIELD SERVICE ENGINEER TIGHTENED THE OIL RETURN VALVE. BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), AND SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). THE DHR WAS REVIEWED. THE UNIT MET MANUFACTURING SPECIFICATIONS AT THE TIME OF RELEASE AND THERE WERE NO ISSUES RELATED TO THIS FAILURE MODE NOTED. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS ((B)(6) 2012 THROUGH (B)(6) 2013) DID NOT OBSERVE ANY SIGNIFICANT TREND. THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS WAS COMPLETED FROM (B)(4) 2013 THROUGH (B)(4) 2013 AND REVEALED A SIGNIFICANT TREND WHICH IS ADDRESSED IN CAPA. THE FMEA REVEALED THE RISK PRIORITY NUMBER (RPN) SCORES ARE BELOW 100 AND IS CONSIDERED ACCEPTABLE. THE SHUMA INDICATES THE RISK IS BROADLY ACCEPTABLE FOR MODERATE LEVEL INJURY AND AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS. CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. THE VACUUM PUMP WAS RETURNED AND TESTED. THE VACUUM PUMP WAS UNABLE TO PASS A PUMP DOWN TEST. THE REASON FOR RETURN OF THE VACUUM PUMP WAS CONFIRMED. THE REPORTED ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.
A CUSTOMER REPORTED AN EVENT OF A "VERY STRONG H2O2" ODOR EMITTING FROM THE STERRAD NX STERILIZER AFTER A LOAD FINISHES. THE CUSTOMER STATES THE ODOR LINGERS FOR ABOUT 5 MINUTES AFTER THE UNIT'S DOOR IS CLOSED. THE ROOM IS VENTILATED AND THERE IS NO MIST/HAZE. THERE WAS NO REPORT OF ANY HUMAN REACTION. THE UNIT WAS TURNED OFF AND AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.
CORRECTED INFORMATION: THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314190 | STERRAD NX STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |