MARYLAND BIPOLAR FORCEPS INSTRRUMENT
Report
- Report Number
- 2955842-2013-02477
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 11, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THERE WAS A MISSING BIPOLAR INSERT FROM THE BACK END OF THE INSTRUMENT, LEAVING THE PINS AND CONDUCTOR WIRES STICKING OUT OF THE CHASSIS. ADDITIONAL OBSERVATION NOT REPORTED BY THE CUSTOMER WAS MAIN TUBE DAMAGE. THE DISTAL END OF MAIN TUBE HAD TWO SMALL SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. ELECTRICAL CONTINUITY TESTING PASSED. NO OTHER DAMAGE WAS FOUND. EVIDENCE NOT CONCLUSIVE, BUT MAIN TUBE DAMAGE MAY BE DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS -HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE MAIN TUBE DAMAGE FOUND DURING FAILURE ANALYSIS COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF TO REOCCUR.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI S SURGICAL PROCEDURE, THE USER FACILITY IDENTIFIED A LOOSE BIPOLAR CONNECTION PART ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. NO MISSING OR FALLEN PIECES WERE REPORTED. THE INSTRUMENT WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309392 | MARYLAND BIPOLAR FORCEPS INSTRRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10121207 115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYSTEM INSTRUMENTS, ACCESSORIES, ESU |