FDA Adverse Event Malfunction Summary report: N

COULTER® AC*T DIFF 2 ANALYZER

MDR report key: 3212150 · Received July 8, 2013

Report

Report Number
1061932-2013-01332
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
June 13, 2013
Report Date
June 13, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K990352
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER CTS (CUSTOMER TECHNICAL SUPPORT) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING OVER THE PHONE. THE CTS ADVISED THE CUSTOMER TO FLUSH THE PROBE WIPE VACUUM PATHWAY AND REPLACE THE AIR FILTERS. THE CTS INDICATED THAT THE FILTERS WERE REPLACED SIX MONTHS AGO AND WERE DUE FOR REPLACEMENT. THE CUSTOMER REPORTED THAT THE INSTRUMENT WAS STILL HAVING PROBLEMS AND WAS GENERATING VACUUM ERRORS. A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) CONTACTED THE CUSTOMER VIA TELEPHONE AND ASSISTED THE CUSTOMER WITH TROUBLESHOOTING. THE FSE SUSPECTED THAT THE AIR FILTERS THAT THE CUSTOMER HAD REPLACED MAY HAVE GOTTEN WET IN THE PROCESS AND WERE THE PROBLEM. THE FSE INSTRUCTED THE CUSTOMER TO REPLACE THE AIR FILTERS WITH NEW ONES AND THE CUSTOMER HAS NOT REPORTED A RECURRENCE OF THE EVENT AS OF (B)(6) 2013. FAILURE MODE OF THE LEAK IS ATTRIBUTED TO THE AIR FILTERS REQUIRING REPLACEMENT. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A DROP OF DILUENT DRIPPED ON TOP OF A SAMPLE TUBE WHILE RUNNING A PATIENT SAMPLE ON THE COULTER AC*T DIFF 2 ANALYZER . THE CUSTOMER INDICATED THAT A FEW DROPS OF FLUID LEAKED AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, GLASSES, AND A LABORATORY COAT AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311515 COULTER® AC*T DIFF 2 ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1