RESERVOIR 1.8ML
Report
- Report Number
- 3004209178-2013-95493
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 18, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER INSPECTION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS AND STOPPER GROOVE FOR DEFECTS AND NONE WERE FOUND. RESERVOIR CONNECTED AND LOCKED IN PLACE PROPERLY.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
IT WAS REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED WHEN SHE REMOVED THE RESERVOIR. THE BLOOD GLUCOSE READING WAS 462MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311476 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A | H8489571 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |