FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3211969 · Received July 8, 2013

Report

Report Number
3004209178-2013-95487
Event Type
Injury
Date Received
July 8, 2013
Date of Event
June 19, 2013
Report Date
June 19, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. UNABLE TO CONFIRM THE NO DELIVERY ALARM. THE DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED. PLEASE SEE MEDWATCH REPORT # 3004209178-2013-95486.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE, AND THE CALL BECAME DISCONNECTED. ATTEMPTED TO CALL THE CUSTOMER SEVERAL TIMES WITH NO RESULTS. INITIALLY, THE CUSTOMER REPORTED HAVING HIGH BLOOD GLUCOSE OF 229MG/DL, AND HIS BLOOD GLUCOSE WAS TREATED WITH MANUAL INJECTIONS. TROUBLESHOOTING WAS PERFORMED, AND THE CALLER STATED THAT THE INSULIN PUMP ALARMED NO DELIVERY DURING MANUAL PRIME. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME AND THE INSULIN DID EXIT. PERFORMED THE DISPLACEMENT TEST AND PASSED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311343 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 40 YR Hospitalization