FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3211965 · Received July 8, 2013

Report

Report Number
3004209178-2013-95472
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
June 17, 2013
Report Date
June 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

UNABLE TO PRIME DURING PRIME ALARM TEST AND ALARMED DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. MISSING END CAP STICKER, CRACKED RESERVOIR TUBE LIP AND CRACKED CASE NEAR RESERVOIR WINDOW NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED AND THE END CAP STICKER WAS MISSING. ADVISED THE CALLER HAT THE DEVICE WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308568 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 15 YR