FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPSINSTRUMENT

MDR report key: 3211861 · Received July 8, 2013

Report

Report Number
2955842-2013-02461
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
May 30, 2013
Report Date
June 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. ONE GRIP WAS BENT, CAUSING A SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .041 OFFSET AT THE INSTRUMENT'S TIPS. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING. ADDITIONAL DAMAGE FOUND WERE DEEP SCRATCHES ON THE MAIN TUBE. THE DISTAL END OF THE MAIN TUBE HAD SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. ENGINEERING CONCLUDED THE DAMAGE MAY HAVE BEEN DUE TO MISHANDLING. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; TUBE ABRASIONS WITH MATERIAL REMOVAL , FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE TIPS WERE NOT MEETING UP ON A FENESTRATED BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311706 FENESTRATED BIPOLAR FORCEPSINSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M11130222 636

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES ESU