UNICEL® DXC 800 PRO SYNCHRON® ACCESS® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2013-00469
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- June 7, 2013
- Report Date
- June 11, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
QUALITY CONTROL (QC) DATA WAS PROVIDED FOR (B)(4) 2013. QC WAS BEGINNING TO SHOW IMPRECISION ON (B)(4) 2013. CUSTOMER STATED THE QC WAS ACCEPTABLE AFTER CALIBRATION. PATIENT SAMPLE INFORMATION AND METHOD OF ANALYSIS WAS NOT PROVIDED BY THE CUSTOMER. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON MULTIPLE DAYS TO EVALUATE THE INSTRUMENT AND CORRECT THE ISSUE. ON (B)(4) 2013, THE FSE REPLACED THE SAMPLE PROBE AND SYRINGE. ON (B)(4) 2013, THE CUSTOMER CALLED BACK AND STATED THAT THE ISSUE WAS NOT RESOLVED. THE FSE REVISITED THE CUSTOMER'S LABORATORY AND REPLACED THE CHLORIDE AND NA ELECTRODES. ON THE SAME DAY, THE CUSTOMER CONTACTED BEC AGAIN STATING THAT THE ISSUE WAS STILL NOT RESOLVED. THE FSE RETURNED AGAIN ON (B)(4) 2013 AND WAS UNABLE TO FIND A DEFINITIVE FAILURE MODE. THE FSE PROACTIVELY REPLACED THE FLOWCELL AND CARBON BRIDGE. THE FOLLOWING DAY ((B)(4) 2013), BEC HOTLINE FOLLOWED UP WITH THE CUSTOMER AND THE CUSTOMER REPORTED THAT THE ISSUE WAS STILL NOT RESOLVED. THE FSE RETURNED AND WAS FINALLY ABLE TO REPLICATE THE EVENT. THE HIGH NA OCCURRED ON THE FIRST SAMPLE AFTER A PERIOD OF INACTIVITY. SERVICE REPLACED THE RATIO PUMP WHICH RESOLVED THE ISSUE. AS OF (B)(4) 2013, THERE HAVE BEEN NO FURTHER IONIZED SELECTIVE ELECTRODE (ISE) ISSUES REPORTED BY THE CUSTOMER. FAILURE MODE WAS ASSOCIATED WITH THE RATIO PUMP. THE FOLLOWING MDRS ARE ASSOCIATED WITH THIS REPORTED EVENT: 2050012-2013-00470, 2050012-2013-00471, 2050012-2013-00472, 2050012-2013-00473.
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER GENERATED FALSE HIGH SODIUM (NA) RESULTS FOR SEVERAL PATIENT SAMPLES ON MULTIPLE DAYS ((B)(6) 2013). THIS REPORT DOCUMENTS THE FIVE (5) PATIENT SAMPLES RAN ON (B)(6) 2013. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE SAMPLES WERE RERUN AND AMENDED REPORTS WERE ISSUED. IT IS UNKNOWN WHICH INSTRUMENT GENERATED THE RERUN RESULTS. NA IS THE ONLY ANALYTE BEING QUESTIONED FOR THIS EVENT. PATIENT DEMOGRAPHICS WERE NOT PROVIDED BY THE CUSTOMER (AGE, DATE OF BIRTH, GENDER, WEIGHT, ETC.). THERE WAS NO DEATH, AFFECT TO PATIENT TREATMENT, OR SERIOUS INJURY REQUIRING MEDICAL ATTENTION ASSOCIATED WITH THE ERRONEOUS RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311662 | UNICEL® DXC 800 PRO SYNCHRON® ACCESS® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |