M2A-MAGNUM 42-50MM TAPER INSERT-6
Report
- Report Number
- 0001825034-2013-02534
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- January 16, 2013
- Report Date
- June 11, 2013
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK042037
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- ATTORNEY
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-02533 / 02535).
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.
LEGAL COUNSEL FOR THE PATIENT REPORTS THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTS THAT A RIGHT REVISION PROCEDURE OCCURRED ON (B)(6) 2013 DUE TO PATIENT ALLEGATIONS OF PAIN, DYSFUNCTION, LOSS OF RANGE OF MOTION, AND ELEVATED METAL ION LEVELS. ADDITIONAL INFORMATION IN REVISION OPERATIVE NOTES INDICATE THE SURGEON NOTED METALLOSIS AND THAT A METAL STAINED CYSTIC MASS WAS REMOVED FROM THE PATIENT¿S WOUND. REVISION OPERATIVE NOTES FURTHER INDICATE OSTEOLYSIS OF THE PROXIMAL FEMUR WITH REACTIVE TISSUE AROUND THE UPPER PORTION OF THE FEMORAL COMPONENT. THE ACETABULAR CUP, MODULAR HEAD, AND TAPER INSERT WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
LEGAL COUNSEL FOR THE PATIENT REPORTS THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTS THAT A RIGHT REVISION PROCEDURE OCCURRED ON (B)(6) 2013, DUE TO PATIENT ALLEGATIONS OF PAIN, DYSFUNCTION, LOSS OF RANGE OF MOTION, AND ELEVATED METAL ION LEVELS. NO FURTHER INFORMATION HAS BEEN PROVIDED TO DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311021 | M2A-MAGNUM 42-50MM TAPER INSERT-6 | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 057730 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |