FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 3211709 · Received July 8, 2013

Report

Report Number
9611451-2013-00519
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE TWO COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. ONE CHAMBER, DEVICE 1, WAS FROM LOT 120510, MANUFACTURED 10 MAY 2012, AND THE OTHER CHAMBER, DEVICE 2, WAS FROM LOT 120525, MANUFACTURED 25 MAY 2012. THE CHAMBERS WERE VISUALLY INSPECTED AND CONNECTED TO A WATER BAG TO CHECK FOR LEAK. RESULTS: VISUAL INSPECTION REVEALED THAT IN BOTH CHAMBERS THE FEEDSET TUBE WAS SLIGHTLY PULLED OUT OF THE BAG SPIKE. SUFFICIENT GLUE WAS FOUND AT THE SPIKE AND FEEDSET TUBE CONNECTION, INDICATING THAT THE SPIKE/FEEDSET TUBE CONNECTION SHOULD HAVE FORMED A TIGHT BOND. DEVICE 1 PASSED THE PERFORMANCE TEST, HOWEVER WHEN THE TEST WAS REPEATED WITH THE FEEDSET UNDER TENSION LEAKAGE OCCURRED FROM THE SPIKE/FEEDSET CONNECTION. LEAKAGE OCCURRED AT THE SPIKE/FEEDSET CONNECTION FROM DEVICE 2 DURING THE PERFORMANCE TEST. A LOT CHECK REVEALED NO SIMILAR COMPLAINTS FOR LOT 120510 AND ONE SIMILAR COMPLAINT FOR LOT 132890. CONCLUSION: THIS TYPE OF DAMAGE CAN BE CAUSED BY THE USER MANIPULATING THE FEEDSET TUBE BY PULLING ON THE TUBING ITSELF, RATHER THAN BY GRASPING THE SPIKE. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE IN ANY OF OUR TESTING. (B)(4). DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT WATER LEAKED AT THE CONNECTION OF THE BAG SPIKE AND THE WATER FEED TUBE OF TWO MR290V HUMIDIFICATION CHAMBERS JUST AFTER START OF USE. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308628 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 120510

Patients

Seq Age Sex Outcome Treatment
1