FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3211708 · Received July 8, 2013

Report

Report Number
1416980-2013-17732
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
June 8, 2013
Report Date
June 15, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE USE ERROR, INAPPROPRIATE BYPASS OF THE DRAIN, NOT INCLUDING I-DRAIN. HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE WARNS THAT "BYPASSING THE DRAIN PHASE CAN LEAVE FLUID IN THE PERITONEAL CAVITY AND RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 03:54:37. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 37995ML, INDICATING THE HOME PATIENT (HP) DRAINED 37995ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309400 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE