FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 3211634 · Received July 8, 2013

Report

Report Number
2124215-2013-08741
Event Type
Injury
Date Received
July 8, 2013
Date of Event
March 27, 2013
Report Date
May 7, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DECLARED A YELLOW ALERT ON THE PATIENT'S HOME MONITORING EQUIPMENT STATING THAT ELECTIVE REPLACEMENT INDICATOR (ERI) HAD BEEN DECLARED. THE PATIENT'S HEALTH CARE PROFESSIONAL (HCP) CALLED BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) FOR CONFIRMATION. TS DISCUSSED ERI INDICATIONS, AND CONFIRMED THAT THE DEVICE DECLARED ERI BASED ON EXTENDED CHARGE TIMES. THE DEVICE WAS EXPLANTED AND REPLACED. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310992 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| L| R 0181| T177