INSIGNIA
Report
- Report Number
- 2124215-2013-07650
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- September 6, 2011
- Report Date
- May 17, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. ON VISUAL INSPECTION, SEVERAL ELECTROCAUTERY MARKS WERE NOTED ON THE PACEMAKER CASE AND ON THE HEADER. THE DEVICE HAD NO TELEMETRY TRANSMISSIONS AND WAS NOT ABLE TO COMMUNICATE WITH SOME OF THE EVALUATION EQUIPMENT. INITIALLY THE PACEMAKER HAD PACING OUTPUTS AT THE RESET-VVI PARAMETERS, BUT THE PACING AMPLITUDE KEPT LOWERING UNTIL THERE WAS NO PACING OUTPUT. THE PACING OUTPUT STOPPED BEFORE THE SENSITIVITY MEASUREMENTS COULD BE TAKEN. THE DEVICE CASE WAS REMOVED AND A PORTION OF THE SENSITIVITY FILTER CIRCUIT CAME LOOSE AND FELL OFF THE INTERNAL CIRCUITRY DURING VISUAL INSPECTION; THIS WAS DUE TO AN OPEN SOLDER JOINT. THIS FINDING COULD CAUSE NOISY SIGNALS AND OVERSENSING AS NOTED IN THE FIELD OBSERVATIONS. THE CAUSE OF THE NO OUTPUT AND NO TELEMETRY WAS ISOLATED TO THE DIGITAL POWER SUPPLY WHICH WAS FOUND TO BE INTERMITTENT WHILE FLEXING THE INTERNAL CIRCUITRY, AND IT WAS DETERMINED THAT THIS WAS INDUCED DAMAGE.
THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER AND RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE AND OVERSENSING WHICH LEAD TO FALSE ATRIAL TACHY RESPONSE (ATR) EPISODES BEING DETECTED. THE DEVICE WAS PROGRAMMED AT DDD 40, AND THE ATRIAL LEAD HAD BEEN PROGRAMMED TO UNIPOLAR SENSING. THE PATIENT WAS NEW TO THE CLINIC SO THE HISTORICAL TRENDS OF THE LEAD AND INDICATION FOR THE PROGRAMMING AND SENSING CONFIGURATION WAS NOT KNOWN. THE HEALTH CARE PROFESSIONAL (HCP) CONSULTED WITH BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) AND PROGRAMMING AND TROUBLESHOOTING OPTIONS WERE DISCUSSED. THE PATIENT HAD REPORTED FEELING LIGHTHEADED AND DIZZY AT TIMES, AND RATES BELOW 40 WERE OCCASIONALLY NOTED DUE TO THE MODIFIED ATRIAL TIMING OF THE EVENTS. ADDITIONAL INFORMATION WAS RECEIVED APPROXIMATELY ONE AND A HALF YEARS AFTER THE PREVIOUS OBSERVATION THAT THE PACEMAKER HAD BEEN EXPLANTED FOR AN UNKNOWN REASON. THERE WERE NO FURTHER ALLEGATIONS OR OBSERVATIONS FROM THE FIELD RELATED TO THE PERFORMANCE OR FUNCTIONALITY OF THIS SYSTEM. THE GENERATOR WAS RETURNED TO BOSTON SCIENTIFIC AND UPON INITIAL EVALUATION THERE APPEARED TO BE A PRODUCT PERFORMANCE ISSUE, AND ADDITIONAL TESTING WAS GOING TO BE PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310394 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | 4469| 4470| 1298| 4471 |