FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 3211588 · Received July 8, 2013

Report

Report Number
2124215-2013-09122
Event Type
Injury
Date Received
July 8, 2013
Date of Event
May 7, 2013
Report Date
May 7, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. VISUAL INSPECTION OF THE ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD BECAME DISLODGED DUE TO TWIDDLER'S SYNDROME. THERE WAS LOSS OF BI-VENTRICULAR PACING DUE TO THE DISLODGEMENT. SURGICAL INTERVENTION WAS PERFORMED WHERE THIS LEAD WAS EXPLANTED AND A NEW LV LEAD WAS IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309504 ACUITY IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4555

Patients

Seq Age Sex Outcome Treatment
1 50 YR Hospitalization| L| R 4092| N119| 4555| 0158| SESR01