FDA Adverse Event Injury Summary report: N

COGNIS

MDR report key: 3211564 · Received July 8, 2013

Report

Report Number
2124215-2013-09002
Event Type
Injury
Date Received
July 8, 2013
Date of Event
March 13, 2013
Report Date
June 25, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS EXPLANTED AND REPLACED APPROXIMATELY THREE MONTHS LATER AND THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

THE LOCAL AREA SALES REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE DEFIBRILLATION LEAD AND CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 200 OHMS BEGINNING APPROXIMATELY TWO MONTHS AGO. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308782 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 57 YR Hospitalization| L| R 0185| N119| 4543| 4136