FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 3211546 · Received July 8, 2013

Report

Report Number
2124215-2013-08947
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
May 10, 2013
Report Date
June 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

THE EXPLANTED LEAD WAS RETURNED AND IS CURRENTLY IN ANALYSIS. THIS REPORT WILL BE UPDATED ONCE ANALYSIS IS COMPLETED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD DISPLAYED OUT OF RANGE HIGH PACING IMPEDANCES DURING THE ATTEMPTED IMPLANT. THE PHYSICIAN ELECTED NOT TO IMPLANT THIS LEAD. THERE WERE NO ADVERSE PATIENT EFFECTS. A NEW LEAD WAS IMPLANTED SUCCESSFULLY.