GYNECARE MORCELLEX TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-12388
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- June 21, 2013
- Report Date
- June 21, 2013
- Manufacturer
- ETHICON INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) - BLADE WILL NOT ADVANCE. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
(B)(4). THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE DID ADVANCE AND DID NOT SEIZE DURING THE EVALUATION. THE DEVICE OPERATED AS INTENDED DURING EVALUATION.
(B)(4). IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE STARTED STUTTERING AND THE BLADE WOULD NOT EXPOSE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308677 | GYNECARE MORCELLEX TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) | HET | ETHICON INC. | NA | MT217669 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |