LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2013-03451
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- June 19, 2013
- Report Date
- June 28, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY, LOCKED OUT AND THE ORANGE INDICATOR WAS FOUND IN ITS PROPER POSITION. THE DEVICE IS DESIGNED TO LOCK OUT WHEN ALL THE CLIPS HAVE BEEN FIRED. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TESTING COULD BE PERFORMED TO EVALUATE THE INCIDENT REPORTED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE EVENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. ADDITIONAL INFORMATION: MULTIPLE REQUEST FOR RETURN OF DEVICE HAVE BEEN MADE BUT NO DEVICE HAS BEEN RETURNED.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WERE MALFORMED. THE CLAMP DELIVERED WELL THE CLIPS BUT THEY WERE NOT FORMED CORRECTLY. THEY ATTACHED WRONG. THEN THE CLAMP STUCK AND WAS UNUSABLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311679 | LIGAMAX-5MM ENDO CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4CN5M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |