ENDOPATH** ETS45
Report
- Report Number
- 3005075853-2013-03448
- Event Type
- Injury
- Date Received
- July 8, 2013
- Date of Event
- June 9, 2013
- Report Date
- June 13, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED YOKE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT AN ETS45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH NO CARTRIDGE RELOAD PRESENT. NO FUNCTIONAL TEST WAS PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE YOKE WAS FOUND BROKEN WHERE IT GUIDES THE SHORT RACK. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE BECAME BROKEN, IT IS POSSIBLE THAT THE DEVICE WAS ATTEMPTED TO FIRE ON TICKER TISSUE THEN INDICATED OR ATTEMPTED TO FIRE TROUGH A LOCKED RELOAD IN PREVIOUS FIRINGS CAUSING AN INCREASE OF THE INTERNAL FORCES RESULTING IN THE COMPONENT YIELDING. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).
IT WAS REPORTED THAT DURING A LEFT NEPHRECTOMY PROCEDURE, THE SURGEON USED THE DEVICE TO TRANSECT THE RENAL ARTERY AND VEIN. MALFORMED STAPLES FOUND AND THE BLOOD WAS BLEEDING. THE SURGEON CHANGED THE PROCEDURE TO OPEN. 1000CC BLOOD TRANSFUSION WAS TAKEN, AND MORE ANESTHETIC WAS USED. THE PROCEDURE WAS PROLONGED ABOUT 20 MIN. AFTER THE PROCEDURE, THE PATIENT WAS SENT TO ICP FOR OBSERVATION ABOUT 12 HOURS, AND NOW THE PATIENT WAS FINE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310061 | ENDOPATH** ETS45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | J4AW46 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |