STERRAD 200, SINGLE DOOR
Report
- Report Number
- 2084725-2013-00309
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 17, 2013
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K030429
- Removal / Correction Number
- Z-0998-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
NI
ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, HEALTH HAZARD EVALUATION, FAILURE MODE AND EFFECTS ANALYSIS AND CORRECTIVE ACTION AND PREVENTIVE ACTION. THE DHR (DEVICE HISTORY REVIEW) FOR STERRAD® 200 CART/RACK CANNOT BE PERFORMED SINCE INFORMATION REGARDING THE PRODUCT'S SERIAL NUMBER IS UNAVAILABLE. TRENDING ANALYSIS FOR "DAMAGED" ISSUES ASSOCIATED TO THE STERRAD® 200 FROM FEBRUARY, 2013 TO JANUARY, 2014 FOUND NO SIGNIFICANT TREND WAS OBSERVED. HHE (HEALTH HAZARD EVALUATION) WAS COMPLETED FOR THIS ISSUE. THE RISK WAS ASSESSED AS NECESSITATES MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE PERMANENT INJURY FOR THE GENERAL POPULATION. FMEA (FAILURE MODE EFFECTS ANALYSIS) WAS REVIEWED. THE RPN FOR THIS FAILURE MODE IS GREATER THAN THE ACCEPTABLE LIMIT AND IS ADDRESSED THROUGH CAPA. CAPA (CORRECTIVE ACTION AND PREVENTIVE ACTION) HAS BEEN INITIATED TO EVALUATE THE WELD ISSUE AND TO REDUCE THE ASSOCIATED RISK. IT WAS DETERMINED THAT INADEQUATE WELDING ON THE STERRAD® 200 CARRIAGE PUSH BAR MAY LEAD TO DISLODGEMENT AND SUBSEQUENT HUMAN INJURY DURING USE. THE ROOT CAUSE OF THE FAILURE OBSERVED HAS BEEN IDENTIFIED AS AN INADEQUATE MANUFACTURING WELD ON THE STERRAD® 200 CARRIAGE PUSH BAR. THE CUSTOMER COMPLAINT DAMAGE TO THE STERRAD® 200 CARRIAGE PUSH BAR HAS BEEN ATTRIBUTED TO A MANUFACTURING ISSUE. A CUSTOMER LETTER WAS SENT INFORMING STERRAD® 200 SYSTEM CARRIAGE OWNERS OF THIS ISSUE. NO FURTHER ACTION IS REQUIRED AT THIS TIME, HOWEVER THE ISSUE WILL CONTINUE TO BE MONITORED.
A CUSTOMER REPORTED THAT THEIR SYSTEM CARRIAGE PUSH BAR WAS 'BROKEN TO PIECES'. THE PUSH CART IS AN ACCESSORY TO THE STERRAD 200 STERILIZER FOR USE IN THE HEALTH CARE (HC) AND SCIENTIFIC AND INDUSTRIAL (S&I) ENVIRONMENTS. THERE WAS NO REPORTED MALFUNCTION OF THE STERRAD 200 STERILIZER GMP. AT THIS TIME, THERE IS NO REPORT OF ANY INJURY OR HARM. ASP WILL CONTINUE FOLLOW-UP EFFORTS. IN THE EVENT ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED. THIS MEDWATCH REPORT IS SUBMITTED AS A MALFUNCTION REPORT AFTER A SERIOUS INJURY TRIGGERING REPORT OF (B)(6) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311412 | STERRAD 200, SINGLE DOOR | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |