CONTAK RENEWAL
Report
- Report Number
- 2124215-2013-07979
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Date of Event
- May 6, 2013
- Report Date
- May 6, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Removal / Correction Number
- Z-0047-08 TO Z-0053-08
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS RETURNED WITH NO REPORTED PRODUCT PERFORMANCE ISSUES AND NO REPORTED ADVERSE PATIENT EFFECTS. INITIAL ANALYSIS COMPLETED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY IDENTIFIED A PRODUCT PERFORMANCE ISSUE. FURTHER ANALYSIS WILL BE PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310676 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H199 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |