FDA Adverse Event Malfunction Summary report: N

RESOLUTION CLIP CLIPPING DEVICE

MDR report key: 3211221 · Received July 8, 2013

Report

Report Number
3005099803-2013-05543
Event Type
Malfunction
Date Received
July 8, 2013
Date of Event
April 18, 2013
Report Date
June 11, 2013
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MND
PMA / PMN Number
K040148
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PATIENT AGE IS UNKNOWN; HOWEVER, THE PATIENT WAS REPORTED TO BE OVER 18 YEARS OLD. INVESTIGATION RESULT OF CLIP MASHED ONTO THE BUSHING. A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE CLIP ASSEMBLY WAS MASHED ONTO THE BUSHING AND THE CLIP ASSEMBLY COULD NOT BE DEPLOYED. THE PRONGS COULD NOT BE OPENED OR CLOSED BUT WERE NOT LOCKED INTO THE CAPSULE. THERE WAS A KINK IN THE CONTROL WIRE. BASED ON THE EVALUATION CONDUCTED AND THE DETAILS OF THE COMPLAINT, THESE FAILURES WERE LIKELY CAUSED BY HANDLING OF THE DEVICE WITHOUT PATIENT CONTACT DURING UNPACKING/PREPARATION. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS HANDLING DAMAGE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS. A SEARCH OF THE COMPLAINT DATABASE CONFIRMED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED BATCH.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USE DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE IN THE FUNDUS ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, DURING PREPARATION OUTSIDE OF THE PATIENT, THE NURSE CLOSED THE CLIP, BUT COULD NOT REOPEN IT. THEREFORE, THE CLIP WAS NOT USED IN THE PATIENT. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED AS BEING STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309840 RESOLUTION CLIP CLIPPING DEVICE LIGATOR, ESOPHAGEAL MND BOSTON SCIENTIFIC - MARLBOROUGH M00522602 ML000383C3

Patients

Seq Age Sex Outcome Treatment
1