ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09846
- Event Type
- Malfunction
- Date Received
- July 8, 2013
- Report Date
- June 10, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/04/2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE REWIND, LOAD CARTRIDGE, AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS. A FORCE SENSOR CALIBRATION TEST CONFIRMED THAT THE SENSOR WAS WITHIN REQUIRED SPECIFICATIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO OCCLUSIONS OCCURRING. AN OCCLUSION WAS INDUCED AND THE PUMP GAVE THE APPROPRIATE VISUAL AND AUDIBLE "OCCLUSION DETECTED" ALARM. THE ISSUE OF THE ABSENCE OF OCCLUSION ALARMS WAS NOT DUPLICATED DURING THE INVESTIGATION. THERE WAS NO DEFECT FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A OCCLUSION (ABSENCE OF OCCLUSION ALARM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309629 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR |