FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3211135 · Received July 8, 2013

Report

Report Number
2124215-2013-08644
Event Type
Injury
Date Received
July 8, 2013
Date of Event
April 17, 2013
Report Date
May 2, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER DEPENDANT PATIENT EXPERIENCED A GREATER THAN 4 SECOND PAUSE IN PACING WHICH LEAD TO ANTI-TACHYCARDIA PACING (ATP). THERE WAS ALSO NOISE PRESENT ON THE ON THE RIGHT VENTRICULAR (RV) LEAD AND LEFT VENTRICULAR (LV) LEAD, AS WELL AS OVERSENSING ON THE RIGHT ATRIAL (RA) LEAD. THE DEVICE, THE RV AND LV LEAD WERE REMOVED AND REPLACED. A NON BOSTON SCIENTIFIC LEAD WAS USED IN THE LV THAT WOULD NOT FIT IN THE CHRONIC TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310352 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0181

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| L| R 0181| 4555| 4054| N119